SHIRAKAWA OLYMPUS CO., LTD. HALOGEN LIGHT SOURCE Back to Search Results
Model Number CLH-SC
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer reported to olympus, the halogen light source fuse blew and the product became unusable.The issue was found during preparation for use.The fuse was replaced.There was no use if implant equipment and there were no reports of patient harm or injury.
 
Manufacturer Narrative
The device was returned to olympus for evaluation and the customer's allegation was not confirmed.The investigation is ongoing and follow up with the user facility has been performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation it was confirmed the fuse was faulty.However, the cause of the faulty fuse was not established at this time.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HALOGEN LIGHT SOURCE
Type of Device
HALOGEN LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18302484
MDR Text Key330180742
Report Number3002808148-2023-14058
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170024481
UDI-Public04953170024481
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K993041
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLH-SC
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/26/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/26/2023
Initial Date FDA Received12/11/2023
Supplement Dates Manufacturer Received01/05/2024
Supplement Dates FDA Received01/16/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/02/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-