MAUDE Adverse Event Report: PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH INTELLIVUE MX450 PATIENT MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH INTELLIVUE MX450 PATIENT MONITOR Back to Search Results
Model Number 866062
Device Problem Device Fell (4014)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/20/2023
Event Type  malfunction  
Event Description
It was reported that the entire front display is defective due to fall damage.It is unknown if the device fell from a mount or if it was dropped.It is unknown if the device was in clinical use at the time of the event, no adverse event or patient harm was reported.The device was returned for bench repair.The repair completion is pending receipt of a part.
 
Manufacturer Narrative
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.E1 reporting address state: (b)(6).E1 reporting institution phone#: (b)(6).
 
Manufacturer Narrative
Diagnostic/functional testing was performed at the philips authorized repair facility.Results of functional testing indicate that the touch bezel and the lcd display were damaged and must be replaced.The bench repair technician (brt) replaced parts (453564406311- iv2-flex dspl lcd assembly 12 and 453564406281-iv2-flex touch bezel assembly) to resolve the issue.A good faith effort (gfe) conducted to obtain data regarding the fall damaged device.The brt indicates that the issues was caused by the user.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE MX450 PATIENT MONITOR
Type of Device
INTELLIVUE MX450 PATIENT MONITOR
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
7031463203
MDR Report Key18305026
MDR Text Key330429440
Report Number9610816-2023-00652
Device Sequence Number1
Product Code MHX
UDI-Device Identifier00884838038769
UDI-Public00884838038769
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K130849
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number866062
Device Catalogue Number866062
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/20/2023
Initial Date FDA Received12/11/2023
Supplement Dates Manufacturer Received01/01/2024
Supplement Dates FDA Received01/10/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/05/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-