Catalog Number 2C8541
Device Problems Disconnection (1171); Fluid/Blood Leak (1250); Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/07/2023
Event Type  malfunction  
Manufacturer Narrative
G1: device manufacturer address 1: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that an unspecified quantity of clearlink duo-vent continu-flo solution sets disconnected during power injection which resulted in a leak.The sets were difficult to connect, and the disconnection occurred when the pressure was applied by the power injector.This issue was further described as, ¿iv blew at the connection/hub¿ and nurses try very hard to get the iv extension set tubing tightened on the threads but still it happens due to the pressure applied by the power injector¿.To resolve this issue, the set was reconnected and was able to continue.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK DUO-VENT CONTINU-FLO SOLUTION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
see h10
see h10
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18307297
MDR Text Key330832434
Report Number1416980-2023-06454
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00085412071107
UDI-Public(01)00085412071107
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K203609
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2C8541
Device Lot NumberR23G12017
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/14/2023
Initial Date FDA Received12/11/2023
Supplement Dates Manufacturer Received01/12/2024
Supplement Dates FDA Received01/16/2024
Date Device Manufactured07/13/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SALINE SOLUTION.
-
-