BAXTER HEALTHCARE CORPORATION CLEARLINK DUO-VENT CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 2C8541 |
Device Problems
Disconnection (1171); Fluid/Blood Leak (1250); Loose or Intermittent Connection (1371)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/07/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
G1: device manufacturer address 1: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that an unspecified quantity of clearlink duo-vent continu-flo solution sets disconnected during power injection which resulted in a leak.The sets were difficult to connect, and the disconnection occurred when the pressure was applied by the power injector.This issue was further described as, ¿iv blew at the connection/hub¿ and nurses try very hard to get the iv extension set tubing tightened on the threads but still it happens due to the pressure applied by the power injector¿.To resolve this issue, the set was reconnected and was able to continue.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
|
|