All available information was investigated and the reported inability to open the clip and retraction problem of the clip introducer were not confirmed via returned device analysis.However, it was observed that the clip introducer was difficult to remove as it was observed that the clip introducer material was caught by the clip arms.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.The investigation was unable to determine a cause for the reported inability to open the clip.The reported retraction problem/observed difficult to remove the clip introducer appears to be due to the user technique of retracting the device.There is no indication of a product issue with respect to manufacture, design, or labeling.
|
It was reported that a patient presented with grade 4 mixed mitral regurgitation (mr) for a mitraclip procedure.The first clip, an xtw clip was aseptically prepared and tested as per instructions for use (ifu).All functionality was good during testing.However, during the first attempt to establish grasping arm angle within the left atrium (la), the clip failed to open.Several troubleshooting attempts were carried out, but clip could not open.The clip delivery system (cds) was therefore removed.It was noted to be difficult retracting the clip into the clip introducer.Once removed, further tests were performed to try and establish fault.It took several attempts before clip eventually opened.Two new xtw clips and one nt clip were subsequently implanted with no issues.There was no adverse effect to patient or unnecessary delay of procedure.No additional information was provided.
|