Model Number N/A |
Device Problem
Packaging Problem (3007)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/28/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that surgeon found the sterile tray was cracked.Another device was used to complete the surgery.Attempts to obtain additional information have been made; however, no more is available.
|
|
Manufacturer Narrative
|
(b)(4).G2: japan.H3: customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual evaluation of the returned device shows the sterile packaging is damaged with corresponding damage to the inside of the carton.This indicates the device shifted during transit contributing to the reported damage.Sterility has been breached.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.The root cause of the reported event can be attributed to transit damage.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|