Model Number 60-05-60 |
Device Problems
Mechanical Problem (1384); Difficult to Open or Close (2921)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/15/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the s5 erc tubing clamp / 500mm.The incident occurred in uk.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova received report that, the arterial clamp displayed error defective clamp changed out during service.There was no patient involvment.
|
|
Manufacturer Narrative
|
H10: the affected part was returned to livanova deutschland for a detailed investigation.The technician could not reproduce the reported issue.Groove ball bearing were replaced.Subsequent functional verification testing was completed without further issues and the unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
See initial report.
|
|