ABBOTT MEDICAL FLEXNAV DELIVERY SYSTEM; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
|
Back to Search Results |
|
Catalog Number FNAV-DS-LG |
Device Problems
Retraction Problem (1536); Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/17/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that on(b)(6) 2023, a 29mm navitor valve was chosen for implantation utilizing a large flexnav delivery system.The navitor was advanced to the valve.During first deployment, position was too high so it was decided to recapture and reposition.During recapture, there was difficulty retracting the valve in the capsule.Both the deployment and micro adjustment wheel was used but bottomed out and were unsuccessful.Kinked portions of the valve capsule were noticed proximally and distally.The flexnav was straightened and with great caution were able to get most of the navitor valve back into the valve capsule where it was removed from the patient.Although, 5-10% of the valve remained outside the valve capsule.There was no issue with the navitor valve itself and retainer tabs remained in place.A replacement 29mm navitor valve and replacement large flexnav delivery system were used to complete the procedure.There were no reported patient consequences and the patient remained hemodynamically stable throughout the procedure.There was a delay that resulted in no patient effects.The patient is reported to be recovering.
|
|
Manufacturer Narrative
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.Na.
|
|
Manufacturer Narrative
|
An event of difficulty retracting the valve in the capsule and kinking in valve capsule was reported.The investigation at abbott found that the valve capsule was deformed damaged and the inner shaft was noted to contain several kinks throughout which precluded using it to perform functional testing.No other anomalies were noted with the device.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Three images from field appeared to show distal end of delivery system having a valve in loaded position, the valve capsule appeared kinked and deformed.Information from field indicated that the valve position was too high and there was difficulty retracting the valve in the capsule.The cause of the reported retraction problem could not be conclusively determined, however the damage noted to the valve capsule is consistent with difficulty retracting the valve.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
|
Search Alerts/Recalls
|
|
|