AIZU OLYMPUS CO., LTD. VIDEOSCOPE "CYF-VA2", WITH SUCTION, EUROPEAN VERSION; CYSTO-NEPHRO VIDEOSCOPE Back to Search Results
Model Number CYF-VA2
Device Problem Device Reprocessing Problem (1091)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation and the customer's allegation of the device was reprocessed without ethylene oxide cap was confirmed.The device evaluation found the device was not reprocessed per the instruction manual.The bending section cover was broken because the cap was not put in correctly, resulting in air/water leakage.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer reported to olympus, the cysto-nephro videoscope was reprocessed without ethylene oxide cap.The issue was found at reprocessing, there was no patient involvement.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and to correct d9.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause could not be identified although it can presumed that it was due to the misuse of users.The event can be detected/prevented in accordance with the following instructions for use: instructions visera cysto-nephro videoscope olympus cyf type v2 olympus cyf type va2 olympus cyf type v2r chapter 7 cleaning, disinfection, and sterilization procedures olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIDEOSCOPE "CYF-VA2", WITH SUCTION, EUROPEAN VERSION
Type of Device
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18312677
MDR Text Key330466019
Report Number9610595-2023-19374
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170339455
UDI-Public04953170339455
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-VA2
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received12/12/2023
Supplement Dates Manufacturer Received01/29/2024
Supplement Dates FDA Received01/29/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-