• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION PERCUFLEX PLUS; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION PERCUFLEX PLUS; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M0061752630
Device Problems Calcified (1077); Difficult to Remove (1528); Use of Device Problem (1670); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Unspecified Infection (1930)
Event Date 11/01/2023
Event Type  Injury  
Manufacturer Narrative
Block b3: date of event was approximated to november 1, 2023 based on the date the manufacturer became aware of the event.Block e1: initial reporter city: (b)(6).Block h6: imdrf device code a150207 captures the reportable event of stent difficult to remove.Imdrf device code a040501 captures the reportable event of stent calcified.Imdrf patient code e0306 captures the reportable event of sepsis.Imdrf patient code e1906 captures the reportable event of infection.
 
Event Description
It was reported to boston scientific corporation that a percuflex plus ureteral stent was successfully implanted during a procedure.The event date was unknown.During a planned stent removal, after five months of being implanted, it was noted that the stent shows a massive encrustation.They used a holmium laser to remove at least enough encrustation from the bladder loop and it took them four hours to remove the stent using a standard stainless-steel gripper under general anesthesia.The patient experienced infection and was given a medication of antibiotic drugs.
 
Manufacturer Narrative
Block b3: date of event was approximated to november 1, 2023 based on the date the manufacturer became aware of the event.Block e1: initial reporter city: (b)(6).Block h6: imdrf device code a150207 captures the reportable event of stent difficult to remove.Imdrf device code a040501 captures the reportable event of stent calcified.Imdrf patient code e0306 captures the reportable event of sepsis.Imdrf patient code e1906 captures the reportable event of infection.Update to block b5: additional information received on 01dec2023, that the problem happened with a timespan within the last six months.
 
Event Description
It was reported to boston scientific corporation that a percuflex plus ureteral stent was successfully implanted during a procedure.The event date was unknown.During a planned stent removal, after five months of being implanted, it was noted that the stent shows a massive encrustation.They used a holmium laser to remove at least enough encrustation from the bladder loop and it took them four hours to remove the stent using a standard stainless-steel gripper under general anesthesia.The patient experienced infection and was given a medication of antibiotic drugs.Additional information received that the problem happened with a timespan within the last six months.
 
Event Description
It was reported to boston scientific corporation that a percuflex plus ureteral stent was successfully implanted during a procedure.The event date was unknown.During a planned stent removal, after five months of being implanted, it was noted that the stent shows a massive encrustation.They used a holmium laser to remove at least enough encrustation from the bladder loop and it took them four hours to remove the stent using a standard stainless-steel gripper under general anesthesia.The patient experienced infection and was given a medication of antibiotic drugs.Additional information received that the problem happened with a timespan within the last six months.
 
Manufacturer Narrative
Block b3: date of event was approximated to november 1, 2023 based on the date the manufacturer became aware of the event.Block e1: initial reporter city: (b)(6).Block h6: imdrf device code a150207 captures the reportable event of stent difficult to remove.Imdrf device code a040501 captures the reportable event of stent calcified.Imdrf patient code e0306 captures the reportable event of sepsis.Imdrf patient code e1906 captures the reportable event of infection.Update to block b5: additional information received on 01dec2023, that the problem happened with a timespan within the last six months.Update to block g2: austria.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCUFLEX PLUS
Type of Device
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key18315354
MDR Text Key330327133
Report Number2124215-2023-68542
Device Sequence Number1
Product Code FAD
UDI-Device Identifier08714729041184
UDI-Public08714729041184
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
K924608
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM0061752630
Device Catalogue Number175-263
Device Lot Number0030582363
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/17/2023
Initial Date FDA Received12/12/2023
Supplement Dates Manufacturer Received12/01/2023
03/06/2024
Supplement Dates FDA Received12/26/2023
03/28/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/22/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-