For the investigation the logfile was analysed, however only limited information were available as part of the entries were already overwritten.Nevertheless, it was possible to retrace the described power supply failure of the hospital on the date of event.As specified the device switched to battery operation and continued ventilation until the mains voltage was available again.No stop of ventilation was found during logfile analysis.Further the logfile does not indicate any problem regarding the reliability of the back-up batteries.In case of a power supply failure, the device continues operation using the internal backup batteries without limitation of functionality while a power failure alarm is given.The remaining capacity of the batteries is indicated to the user and additional warnings are given when the residual battery charge is low respectively very low.Dräger finally concludes that the device behaved a s specified.
|