• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS TROPONIN T HS STAT; IMMUNOASSAY METHOD, TROPONIN SUBUNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS TROPONIN T HS STAT; IMMUNOASSAY METHOD, TROPONIN SUBUNIT Back to Search Results
Catalog Number 09315349190
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Low Test Results (2458); Non Reproducible Results (4029)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/21/2023
Event Type  malfunction  
Event Description
We received an allegation about discrepant results for 1 patient's sample tested with elecsys troponin t hs (tnths) stat assay on a cobas e411 immunoassay analyzer.Initial result: 27.Repeat result: 2717.The units of measurement were not provided.
 
Manufacturer Narrative
The tnths stat reagent lot number wand expiration date were not provided.The cobas e411 rack serial number was (b)(6).The investigation is ongoing.
 
Manufacturer Narrative
A general reagent problem can be excluded because the qc prior to the event was within ranges.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS TROPONIN T HS STAT
Type of Device
IMMUNOASSAY METHOD, TROPONIN SUBUNIT
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18317273
MDR Text Key330362645
Report Number1823260-2023-03966
Device Sequence Number1
Product Code MMI
Combination Product (y/n)Y
Reporter Country CodeCA
PMA/PMN Number
K201441NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number09315349190
Device Lot Number726126
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/21/2023
Initial Date FDA Received12/13/2023
Supplement Dates Manufacturer Received12/21/2023
Supplement Dates FDA Received12/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-