• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH INTELLIVUE MX450 PATIENT MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH INTELLIVUE MX450 PATIENT MONITOR Back to Search Results
Model Number 866062
Device Problem High Readings (2459)
Patient Problem Insufficient Information (4580)
Event Date 11/15/2023
Event Type  malfunction  
Manufacturer Narrative
Philips is in the process of obtaining additional information and the complaint is still under investigation.A follow-up report will be submitted upon completion of the investigation.Reporter telephone number: (b)(6).
 
Event Description
The customer reported the patient's blood pressure was too high and the patient was given antihypertensive drugs.The blood pressure did not drop and the user replaced the intellivue mx450 patient monitor with another brand and the blood pressure was too low.The device was reported to be in use on a patient, but no adverse event to the patient or user was reported.
 
Manufacturer Narrative
A philips representative spoke by telephone support with the biomed who performed functional testing of the device and identified the nibp cuff was faulty.A replacement nibp cuff was provided to the customer.Additional information received via gfe indicated there was no harm related to the administration of the medication.The investigation concludes that no further action is required at this time.The device remains at the customer site.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE MX450 PATIENT MONITOR
Type of Device
INTELLIVUE MX450 PATIENT MONITOR
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
7031463203
MDR Report Key18318158
MDR Text Key330378664
Report Number9610816-2023-00653
Device Sequence Number1
Product Code MHX
UDI-Device Identifier00884838038769
UDI-Public00884838038769
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K130849
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number866062
Device Catalogue Number866062
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/01/2023
Initial Date FDA Received12/13/2023
Supplement Dates Manufacturer Received01/10/2024
Supplement Dates FDA Received02/09/2024
Date Device Manufactured05/28/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-