MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC CVC SET: 2-LUMEN 7 FR X 20 CM; CATHETER INTRAVASCULAR THERAPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC CVC SET: 2-LUMEN 7 FR X 20 CM; CATHETER INTRAVASCULAR THERAPE Back to Search Results
Catalog Number CS-27702-E
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Respiratory Arrest (4461)
Event Date 11/16/2023
Event Type  Injury  
Manufacturer Narrative
Qn# (b)(4).
 
Event Description
It was reported that: (b)(6) 2023, the doctor found the patient was in shock, respiratory arrest, and emergency resuscitation conditions during catheter use.After that cvc removal and rescue medication treatment.The patient condition is fine.Additional information has been requested from the account.A follow up report will be submitted after investigation.
 
Event Description
It was reported that: (b)(6) 2023, the doctor found the patient was in shock, respiratory arrest, and emergency resuscitation conditions during catheter use.After that cvc removal and rescue medication treatment.The patient condition is fine.
 
Manufacturer Narrative
(b)(4).The complaint of catheter related allergic reaction could not be confirmed.No sample was returned for investigation.The customer did not report that allergy testing was performed to confirm a chlorhexidine allergy.The instructions for use (ifu) provided with this kit warns the user, "the arrowg+ard blue antimicrobial catheter is contraindicated for patients with known hypersensitivity to chlorhexidine, silver sulfadiazine and/or sulfa drugs.Remove catheter immediately if adverse reactions occur after catheter placement.Chlorhexidine containing compounds have been used as topical disinfectants since the mid-1970's.An effective antimicrobial agent, chlorhexidine found use in many antiseptic skin creams, mouth rinses, cosmetic products, medical devices and disinfectants used to prepare the skin for a surgical procedure.Note: perform sensitivity testing to confirm allergy to catheter antimicrobial agents, if adverse reaction occurs." a device history record review was performed, and no relevant findings were identified to suggest a manufacturing related issue.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CVC SET: 2-LUMEN 7 FR X 20 CM
Type of Device
CATHETER INTRAVASCULAR THERAPE
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key18318556
MDR Text Key330383261
Report Number3006425876-2023-01239
Device Sequence Number1
Product Code FOZ
UDI-Device Identifier00801902102362
UDI-Public00801902102362
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K900263
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberCS-27702-E
Device Lot Number71F23A2872
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/16/2023
Initial Date FDA Received12/13/2023
Supplement Dates Manufacturer Received12/21/2023
Supplement Dates FDA Received01/03/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/06/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NOT REPORTED; NOT REPORTED
-
-