• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK IRELAND LTD EVOLUTION® ESOPHAGEAL CONTROLLED-RELEASE STENT - FULLY COVERED; ESW PROSTHESIS, ESOPHAGEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK IRELAND LTD EVOLUTION® ESOPHAGEAL CONTROLLED-RELEASE STENT - FULLY COVERED; ESW PROSTHESIS, ESOPHAGEAL Back to Search Results
Catalog Number EVO-FC-R-18-23-12-E
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/16/2023
Event Type  malfunction  
Manufacturer Narrative
Investigation is still pending, a follow up mdr will be submitted to include the investigation conclusions.
 
Event Description
The stent couldn't be released as it remained stuck in the introducer "as per cc form": after positioning, after the removal maneuver of the prosthesis positioner, we witness the removal of the newly positioned prosthesis, as it remains hooked to the removal wire.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence 1.At what stage of the procedure did the complaint occur? when unpacking or preparing the evolution, while inserting the evolution in the patient, during stent placement, while removing the introducer, or during stent repositioning/removal while removing the introducer.2.What endoscope type and channel size was used? no endoscope 3.What was the position of the elevator? n/a, open, closed no elevator 4.Details of the wire guide used (diameter, type, make)? metii-35-80 5.Was the zip port facing upwards and slightly curved when backloading the wire guide? n/a, yes, no no zip port 6.Did any part of the stent contact the patient¿s anatomy when the complaint occurred? n/a, yes, no esophagus 7.Please advise the anatomical location of the intended target site.Esophagus 8.How long was the stent in the patient by the time this complaint occurred? 9.For devices where the ifu states for longer term patency has not been established, was periodic evaluation completed? n/a, yes, no 10.If yes, how often was this completed? 11.Did the patient require any additional procedures as a result of this event? n/a, yes, no no 12.What intervention (if any) was required? none 13.Was the secondary intervention performed during the same procedure as the device failure or was it scheduled for another day? n/a, same procedure, another day 14.Were any other defects (other than the complaint issue) observed on the device prior to return (e.G.Kink)? n/a, yes, no no 15.If yes, please specify what was observed and where on the device it was observed.Stricture information: 1.What was the length and diameter of the stricture? 2.Where was the stricture located in the body? 3.Was there resistance felt passing wire guide through stricture? n/a, yes, no no 4.Was there resistance felt passing the evolution through stricture? n/a, yes, no no 5.Was the stricture dilated before stent placement? n/a yes, no no questions related to during insertion into patient 1.Was the product inspected for kinks or damage before use? n/a, yes, no 2.Was resistance felt during insertion into patient? n/a, yes, no 3.If yes, at what point? questions related to during stent placement 1.Did the product fail during stent deployment or recapture? n/a, deployment, recapture, other during stent deployment 2.If other, please specify 3.Was the directional button pressed during use? n/a, yes, no yes 4.Was any part of the stent observed in contact with the patient¿s anatomy at the time of failure? n/a, yes, no 5.Was the yellow marker kept in view during deployment? n/a, yes, no 6.Are images of the device or procedure available? n/a, yes, no] yes questions related to during introducer withdrawal 1.Are images of the device or procedure available? n/a, yes, no yes 2.Was final stent placement confirmed using endoscopy / fluoroscopy? n/a, yes, no fluoroscopy 3.If yes, what was used? 4.Did the stent open sufficiently to allow withdrawal of introducer safely? n/a, yes, no no 5.Was the safety wire fully removed before removing the delivery system? n/a, yes, no yes 6.Did any part of the product snag/get caught with the stent when removing the delivery system? n/a, yes, no loop removing delivery system questions related to during stent repositioning/removal (for evo-fc & evo-pc devices) 1.What instrument was used for stent repositioning / removal? forceps, snare, other 2.If other, please specify.3.Was resistance encountered during advancement and/or deployment? n/a, yes, no 4.If yes, please when this was felt? advancement or deployment 5.How did the physician deal with this resistance? 6.Was the lasso (suture) loop used during repositioning patient not affected by covid.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUTION® ESOPHAGEAL CONTROLLED-RELEASE STENT - FULLY COVERED
Type of Device
ESW PROSTHESIS, ESOPHAGEAL
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
Manufacturer (Section G)
COOK IRELAND LTD
o halloran road
national technology park
limerick
Manufacturer Contact
sinead o'leary
o halloran road
national technology park
limerick 
MDR Report Key18319076
MDR Text Key330663615
Report Number3001845648-2023-00901
Device Sequence Number1
Product Code ESW
UDI-Device Identifier10827002551692
UDI-Public(01)10827002551692(17)250619(10)C2062662
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K093619
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEVO-FC-R-18-23-12-E
Device Lot NumberC2062662
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/29/2023
Was the Report Sent to FDA? No
Distributor Facility Aware Date11/16/2023
Event Location Hospital
Initial Date Manufacturer Received 11/16/2023
Initial Date FDA Received12/13/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/19/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age66 YR
Patient SexMale
-
-