The joimax tessys device still resides at the user facility to the manufacturer's knowledge.No device identifiers are available at this time.Approximately one year ago (october 17, 2022), two joimax representatives visited the hospital to perform preventive maintenance; during this visit, extensive rust was observed on the ilessys and tessys instrument sets and joimax identified that the devices had been improperly reprocessed.The joimax representatives informed the hospital of these findings and offered to provide retraining in accordance with the device labeling.The hospital declined the retraining.The device user manual and reprocessing instructions clearly state that after cleaning and disinfection, all medical devices must be inspected for function, corrosion, damaged surfaces, chipping or / and cleanliness.Specifically, "if a deterioration / defect is determined, then the instrument is to be segregated immediately and may not be used further.Inadequate or omitted reprocessing can lead to infection or inflammation of the patient".Based on the information reviewed, the root cause of the infection was likely attributed to improper device reprocessing (use error).Infection is identified in the system risk assessment.Manufacturer reference #: (b)(4).
|