MEDTRONIC PUERTO RICO OPERATIONS CO. INTERSTIM X; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON
|
Back to Search Results |
|
Model Number 97800 |
Device Problems
Use of Incorrect Control/Treatment Settings (1126); Display or Visual Feedback Problem (1184)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/15/2023 |
Event Type
malfunction
|
Event Description
|
Information was received from a manufacturer representative regarding a patient who was implanted with an implantable neurostimulator (ins) for unknown indications for use.It was reported that patient with a sacral nerve stimulator attended their mri scan.The patient programmer was used to identify if they were eligible.The programmer said they weren't.The consultant and manufacturer was consulted and it was found that the patient devices was incorrectly programmed with the wrong lead model number.The patient should have been eligible for their mri.The device was re-programmed.The correct lead details added onto programmer, so patient was able to have mri.Issue resolved.
|
|
Manufacturer Narrative
|
G2: country united kingdom medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received.The manufacturer representative (rep) reported the indication for use was overactive bladder.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|