Catalog Number 451570101 |
Device Problem
Unintended System Motion (1430)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/04/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during a total knee arthroplasty surgical procedure, it was observed that while using the robotic-assisted solution satellite station device it was noticed that the while making the anterior chamfer cut, the robot made a sudden move in the opposite direction.Then they were prompted to move the device parallel to the leg and the case was completed with the robot.It was reported that nothing moved.All checkpoints were verified to be as originally registered.The surgeon was happy with the results.It was reported that the knee looked and felt great.There were no gaps.However on the final leg alignment screen the gaps were huge and way off from original plan.There were two cases following this one without issue.The reporter suspects something moved to cause the robot to behave this way.It was reported that the device was being used with a robotic assisted base station device.There were no delays in the procedure.There was patient involvement.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.D10, concomitant medical devices and therapy dates, base station device, december 12, 2023.Udi: (b)(4).
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The device log file for this event was reviewed.The investigation could not confirm sudden robot movement.The investigation revealed visible leg motion during the anterior chamfer resection, emphasizing the recommendation to stabilize the leg before each resection to ensure accuracy.The issue related to the prompt to move device parallel to leg could not be confirmed since the logs did not contain the message "align device array parallel to knee" therefore this issue could not be confirmed.The review was unable to confirm the report of gaps since the the user did not use the pointer tool to verify any cuts, so the investigation cannot confirm the alleged issue.There were no defects found with the system and software.No problem was found.
|
|
Search Alerts/Recalls
|
|