MAUDE Adverse Event Report: ST. JUDE MEDICAL TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿ BID CURVE F-J; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿ BID CURVE F-J; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number A-TCSE-FJ
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Cardiac Perforation (2513)
Event Date 12/06/2023
Event Type  Injury  
Event Description
During a left premature ventricular contractions ablation procedure, a pericardial effusion occurred which required a pericardiocentesis to stabilize the patient.During ablation, high co2 level and hypotension were noted.The procedure was stopped and an echocardiogram was performed which confirmed an effusion, the perforation was located at mid-anterior side of the left ventricle.A pericardiocentesis was done to stabilize the patient.There were no reported performance issues with any abbott device.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis; however, the log files from the tactisys quartz system were returned.The log file analysis concluded that the tacticath catheter performed as intended.The optical fibers met specifications, the recorded temperatures indicated cooling during rf ablation, and contact force measurements were displayed throughout the duration of the log files.Due to unknown procedural conditions, we are unable to conclusively determine the cause of the reported event.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Per the ifu, cardiac tamponade and pericardial effusion are known risks during the use of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TACTICATH¿ CONTACT FORCE ABLATION CATHETER, SENSOR ENABLED¿ BID CURVE F-J
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key18325036
MDR Text Key330486418
Report Number3008452825-2023-00565
Device Sequence Number1
Product Code OAE
UDI-Device Identifier05415067027665
UDI-Public05415067027665
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P130026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberA-TCSE-FJ
Device Lot Number9253689
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/06/2023
Initial Date FDA Received12/14/2023
Supplement Dates Manufacturer Received12/22/2023
Supplement Dates FDA Received01/02/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/21/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Life Threatening;
-
-