Model Number CARDIOHELP.I
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/04/2023
Event Type  malfunction  
Manufacturer Narrative
It was reported that the venous bubble sensor was alarming.It was later specified, that the device showed the error message "venous bubble sensor is defective".Cleaning the sensor elevated the problem for a short period of time, but it reoccurred.The failure occurred during treatment.A getinge field service technician (fst) was sent for investigation and repair.The fst could replicate the failure by moving the cable of the sensor and eventually, the venous bubble sensor was replaced.The fst performed safety, calibration, and functionality checks to factory specifications.All function tests are passed.The log files of the reported cardiohelp device were reviewed and the error message "venous bubble sensor is defective" could be confirmed on the date of event.No exact root cause could be determined.However, another venous bubble sensor with a similar failure was already investigated by the supplier sonotec.Following possible root causes were determined: damaged wiring inside the cable due to mechanical tension - damage due to overvoltage or esd (electrostatic discharge) according to the instructions for use (ifu), chapters 2.2.5 "monitoring and sensors" and 5.4.4 "bubble monitoring: function test" the venous bubble sensor and the arterial flow/bubble sensor have to be tested before each use.The cardiohelp has a flow/bubble sensor for bubble detection.The venous bubble sensor is optional and for additional bubble detection.The review of the non-conformities has been performed on 2023-12-13 for the period of 2017-04-26 to 2023-12-04.It does not show any non-conformity in regard to the reported product and failure.There is no indication that manufacturing issues occurred during this time, thus production related influences are unlikely.The device was manufactured on 2017-04-26.In order to avoid reoccurrence of the reported failure, the customer will be informed by the getinge sales and service unit (ssu) to follow the chapter 5.3."connecting the sensors" in the instructions for use of the cardiohelp device.Based on the results the reported failure "error message: venous bubble sensor" could be confirmed.The customer will be informed about the results by the getinge sales and service unit.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿ s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
It was reported that the venous bubble sensor was alarming ¿venous bubble sensor is defective¿ when moving the cable.No harm to any person has been reported.Complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOHELP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key18325069
MDR Text Key330486566
Report Number8010762-2023-00619
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCARDIOHELP.I
Device Catalogue Number701048012
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/04/2023
Initial Date FDA Received12/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/26/2017
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-