OLYMPUS WINTER & IBE GMBH RESECTION SHEATH, 8 MM, FOR 8.5 MM/26 FR. OUTER SHEATH, ABS; RESECTOSCOPE
|
Back to Search Results |
|
Model Number A42011A |
Device Problems
Break (1069); Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/27/2023 |
Event Type
Injury
|
Event Description
|
The customer reported to olympus that the ceramic tip of the resection sheath, 8 mm, for 8.5 mm/26 fr.Outer sheath, abs broke off during a cystoscopy with laser usage procedure.The tip was removed from the patient without any problem or damage to the patient.The customer used the resection sheath in combination with an a22021 (outer sheath, 27 fr., 2 stopcocks, rotatable), which is the wrong combination according to the territory manager.There was no further patient impact reported.
|
|
Manufacturer Narrative
|
The suspect device referenced in this report has not been returned to olympus.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if additional information is received.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is possible the ceramic tip broke off and was retrieved occurred due to combination of incompatible articles.However, the subject device was not returned and the root cause could not be identified.The event can be prevented by following the instructions for use which state: -warning -infection control risk "properly reprocess the product before first and each subsequent use following the instructions in this manual and in the system guide endoscopy.Improper and/or incomplete reprocessing can cause infection of the patient and/or medical personnel." -4.1 inspection and testing -inspecting the product "visually inspect the product.Make sure that it has: -- no corrosion -- no dents -- no scratches ceramic insulation at distal end visually inspect the ceramic insulation at the sheath¿s distal end before each use.Do not use the instrument in case of damage (e.G.Cracks, fractures)." -warning -risk of injury "impact, fall, shock or similar stress can damage the ceramic insulation at the sheath¿s distal end.Damaged instruments can cause injuries to the patient and/or user.Do not use the instrument if damaged." olympus will continue to monitor field performance for this device.
|
|