Updated: h4, h6, h10 this report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the reported broken cutting loop wire was confirmed.A root cause could not be determined, however, the issue was likely due to user handling.Olympus will continue to monitor field performance for this device.
|