Model Number FRX1130S14
Device Problems Corroded (1131); Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2023
Event Type  malfunction  
Event Description
The manufacturer received information regarding a dreamstation avaps30.The device was returned to a third party service center.During visual inspection of the device, corrosion was found on the power connector and the unit can't be power on.Dust/dirt contamination was also found in the device.This report is being submitted for the corrosion found on the power connector during service.There was no allegation of serious or permanent harm or injury.No medical intervention was required for the patient.Device was scrapped.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Manufacturer Narrative
The manufacturer received information in relation to a dreamstation avaps unit.The device was returned to a third party service center.During visual inspection of the device, it was determined the power connector of the unit was corroded.Device was scrapped at customer's request.Analysis of past events has not indicated an adverse event has occurred due to corrosion.Review of the risk file indicates the potential for a serious adverse event occurring as a result of this incident is unlikely.Additionally, the risk file indicates that corrosion will not substantially affect the performance of the device.This complaint is considered not reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AVAPS30
Type of Device
VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key18326428
MDR Text Key330497623
Report Number2518422-2023-35552
Device Sequence Number1
Product Code MNS
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K102465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberFRX1130S14
Device Catalogue NumberFRX1130S14
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/01/2023
Initial Date FDA Received12/14/2023
Supplement Dates Manufacturer Received12/01/2023
Supplement Dates FDA Received02/05/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured02/08/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-