ROCHE DIAGNOSTICS COAGUCHEK XS PT TEST PST; PROTHROMBIN TIME MONITOR Back to Search Results
Catalog Number 07671687016
Device Problem High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/17/2023
Event Type  malfunction  
Event Description
We received an allegation of discrepant inr results with a coaguchek inrange meter compared to the werfen laboratory method.The result from the meter was 2.6 inr.The result from the laboratory within 3 hours was 1.8 inr.The patient¿s therapeutic range is 2 ¿ 3 inr.
 
Manufacturer Narrative
Section e3: occupation is patient/consumer (patient¿s mother).The coaguchek inrange meter serial number was (b)(6).On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.Per product labeling, "coaguchek uses human recombinant thromboplastin.Therefore, the comparability to other human recombinant thromboplastins is best, whereas higher deviations can occur with other thromboplastins.However, those higher differences between thromboplastins of different (rabbit, bovine based) origin are not a coaguchek specific issue.Similar differences can be observed when a human recombinant thromboplastin-based laboratory method is compared against several other (rabbit, bovine-based) laboratory methods.".
 
Manufacturer Narrative
Section h6, additional type of investigation code provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK XS PT TEST PST
Type of Device
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18326520
MDR Text Key330498337
Report Number1823260-2023-03994
Device Sequence Number1
Product Code GJS
Combination Product (y/n)Y
Reporter Country CodeFR
PMA/PMN Number
K170960
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number07671687016
Device Lot Number73986113
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 11/22/2023
Initial Date FDA Received12/14/2023
Supplement Dates Manufacturer Received01/12/2024
Supplement Dates FDA Received01/16/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
WARFARIN.
Patient Age20 YR
Patient SexMale
-
-