• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH INTELLIVUE MULTI MEASUREMENT SERVER X2

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH INTELLIVUE MULTI MEASUREMENT SERVER X2 Back to Search Results
Model Number M3002A
Device Problems Inappropriate Audible Prompt/Feedback (2280); No Audible Prompt/Feedback (2282)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/04/2023
Event Type  malfunction  
Manufacturer Narrative
A follow up report will be submitted upon completion of the investigation.E1: reporting institution phone: (b)(6).E1: reporter phone: (b)(6).
 
Event Description
It was reported the intellivue multi measurement server x2 speaker is defective and needed replacement.It is unknown if there was still sound.A loss of audio cannot be ruled out based on information currently available.The device was in use on a patient.There was a report of patient harm.
 
Manufacturer Narrative
A remote service engineer (rse) determined the speaker required replacement.Based on the information available and the testing conducted, the cause of the reported problem was the speaker.The reported problem was confirmed.The customer was provided a replacement speaker to resolve the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE MULTI MEASUREMENT SERVER X2
Type of Device
INTELLIVUE MULTI MEASUREMENT SERVER X2
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
7031463203
MDR Report Key18333893
MDR Text Key330554864
Report Number9610816-2023-00663
Device Sequence Number1
Product Code DSI
UDI-Device Identifier00884838000261
UDI-Public00884838000261
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K150310
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM3002A
Device Catalogue NumberM3002A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/04/2023
Initial Date FDA Received12/15/2023
Supplement Dates Manufacturer Received01/08/2024
Supplement Dates FDA Received02/01/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/02/2014
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-