• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS SYPHILIS; ENZYME LINKED IMMUNOABSORPTION ASSAY, TREPONEMA PALLIDUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS SYPHILIS; ENZYME LINKED IMMUNOABSORPTION ASSAY, TREPONEMA PALLIDUM Back to Search Results
Catalog Number 07251378190
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/24/2023
Event Type  malfunction  
Manufacturer Narrative
The cobas e 801 analytical unit serial number was (b)(6).The investigation is ongoing.
 
Event Description
There was an allegation of a questionable syphilis elecsys e2g result from the cobas e 801 analytical unit.The result from the cobas e801 was 0.0892 (negative).The result by trust method (shanghai rongsheng) was weak positive 1:1.At another site, the rpr result was weak positive 1:1.
 
Manufacturer Narrative
The investigation determined the issue was consistent with a sample-specific discrepancy.The elecsys syphilis assay performed within specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS SYPHILIS
Type of Device
ENZYME LINKED IMMUNOABSORPTION ASSAY, TREPONEMA PALLIDUM
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250
MDR Report Key18335098
MDR Text Key331050993
Report Number1823260-2023-04027
Device Sequence Number1
Product Code LIP
UDI-Device Identifier04015630940356
UDI-Public04015630940356
Combination Product (y/n)Y
Reporter Country CodeCH
PMA/PMN Number
K160910
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2023
Device Catalogue Number07251378190
Device Lot Number682565
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/28/2023
Initial Date FDA Received12/15/2023
Supplement Dates Manufacturer Received12/22/2023
Supplement Dates FDA Received12/22/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-