Catalog Number 07251378190 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/24/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The cobas e 801 analytical unit serial number was (b)(6).The investigation is ongoing.
|
|
Event Description
|
There was an allegation of a questionable syphilis elecsys e2g result from the cobas e 801 analytical unit.The result from the cobas e801 was 0.0892 (negative).The result by trust method (shanghai rongsheng) was weak positive 1:1.At another site, the rpr result was weak positive 1:1.
|
|
Manufacturer Narrative
|
The investigation determined the issue was consistent with a sample-specific discrepancy.The elecsys syphilis assay performed within specification.
|
|