• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR PRESSUREWIRE¿ X GUIDEWIRE; CATHETER TIP PRESSURE TRANSDUCER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR PRESSUREWIRE¿ X GUIDEWIRE; CATHETER TIP PRESSURE TRANSDUCER Back to Search Results
Catalog Number C12059
Device Problems Material Separation (1562); Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/28/2023
Event Type  Injury  
Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Event Description
It was reported that during the procedure, the pressurewire x wireless device was to be used in the mildly calcified, mildly tortuous, left anterior descending (lad) artery.The device was calibrated and equalized without issue.The device was being advanced with resistance, when the tip of the device separated inside of the anatomy.A snare device was used to remove the separated tip, and an unspecified device was used successfully for the procedure.There was no adverse patient sequela and no reported clinically significant delay in the procedure or therapy.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no indication of a lot specific product quality issue.Based on the reported information, the investigation determined that the reported failure to advance resulting in material separation and unexpected medical intervention was likely due to circumstances of the procedure.It is likely that during the failed attempt to advance, the tip of the pressurewire was subjected to tensile or torsional loads beyond its design limits causing the guide wire to separate.Anatomical conditions likely contributed to the failure to advance.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSUREWIRE¿ X GUIDEWIRE
Type of Device
CATHETER TIP PRESSURE TRANSDUCER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key18340573
MDR Text Key330660261
Report Number2024168-2023-14112
Device Sequence Number1
Product Code DXO
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K180558
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberC12059
Device Lot Number30428G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/07/2023
Initial Date FDA Received12/15/2023
Supplement Dates Manufacturer Received01/17/2024
Supplement Dates FDA Received02/08/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/28/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-