Model Number 3300TFX25MM
Device Problem Perivalvular Leak (1457)
Patient Problem Dyspnea (1816)
Event Date 11/22/2023
Event Type  Injury  
Manufacturer Narrative
H10: additional manufacturer narrative: the investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event Description
It was reported that a patient with a 25mm 3300tfx pericardial aortic valve underwent a valve-in-valve procedure after an approximate implant duration of nine (9) years due to pvl.The patient initially presented with shortness of breath.The tavr procedure was performed with a 26mm 9750tfx transcatheter valve.The patient was noted to be stable after the tavr procedure.Per the team, there was no allegation of any edwards device malfunction or premature failure.
 
Manufacturer Narrative
The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.Regurgitation is considered to be a perivalvular leak (pvl) if a turbulent eccentric jet originates between the bioprosthetic sewing ring and the annulus.Pvl can occur in the mitral and aortic position for similar reasons.In the early postoperative period, the highest incidence of pvl has been seen in patients developing infective endocarditis, which is most likely attributed to inadequate peri-operative antibiotic prophylaxis or nosocomial infection.Annular calcification is also a risk factor for the development of peri-operative pvl as the bioprosthesis may not seat properly after debridement.Technique related factors, such as incorrect valve sizing, have been shown to contribute to the development of pvl.Anatomical factors may create difficulty seating the bioprosthetic valve resulting in pvl.The anatomy of the annulus may induce mechanical stresses along the rigid bioprosthetic ring which can influence long-term valve performance and durability.A diseased or rigid annulus can potentially increase the mechanical stress on the prosthetic valve, leading to pvl.An engineering evaluation is not required because there is no allegation of a malfunction which could be related to a manufacturing non-conformance and/or one was not suspected or confirmed through investigation; no labeling non-conformance/deficiency; no use-related issue with a hazardous situation; no device-related infection; and no evidence of a product failure with regard to design, reliability, or use error.The instructions for use (ifu) have been reviewed and no inadequacies have been identified with regard to warnings, contraindications, and the directions/conditions for the successful use of the device.The reported type of event is included in the ifu.A capa/scar/pra is not required as there are no confirmed product or labeling non-conformances and no other triggers are met.A definitive root cause cannot be conclusively determined; however, patient and/or procedural factors likely caused or contributed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT MAGNA EASE PERICARDIAL BIOPROSTHESIS
Type of Device
REPLACEMENT HEART-VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
irvine, CA 92614
9492506615
MDR Report Key18340984
MDR Text Key330663909
Report Number2015691-2023-18445
Device Sequence Number1
Product Code DYE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Expiration Date05/17/2018
Device Model Number3300TFX25MM
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/22/2023
Initial Date FDA Received12/15/2023
Supplement Dates Manufacturer Received01/31/2024
Supplement Dates FDA Received02/05/2024
Date Device Manufactured05/18/2014
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Life Threatening; Hospitalization;
Patient Age57 YR
Patient SexMale
-
-