RESPIRONICS, INC. DREAMSTATION AUTO; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number FRX500S14
Device Problems Contamination (1120); Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/18/2023
Event Type  malfunction  
Manufacturer Narrative
H3 other text : device serviced by third party service center.
 
Event Description
The manufacturer received information regarding a dreamstation auto.The device was returned to a third party service center.During visual inspection of the device, the power connector of the unit or the electronic board seems defective, and the unit cannot be powered on in order to collect the error log, error numbers, blower and machine hours, software version.In addition, the inspection found dust and dirt were observed inside the device.The device was scrapped.This report is being submitted for the power connector of the unit or electronic board seems defective, and the unit cannot be powered on.There was no allegation of serious or permanent harm or injury.No medical intervention was required for the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AUTO
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key18343065
MDR Text Key330783308
Report Number2518422-2023-36079
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K131982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberFRX500S14
Device Catalogue NumberFRX500S14
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/18/2023
Initial Date FDA Received12/18/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/24/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-