Model Number CLV-190
Device Problem Moisture Damage (1405)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/23/2023
Event Type  malfunction  
Event Description
The customer reported to olympus, the xenon light source had water entering the compressed air system.There were no reports of patient harm.
 
Manufacturer Narrative
The device was not returned to olympus for evaluation.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over nine (9) years since the subject device was manufactured.During the inspection, olympus confirmed the reported event: water was leaking from the output connector socket, causing flooding inside the device and liquid adherence to the board.In addition, the following non-reportable malfunctions were found during the device evaluation: a damaged front panel and dust inside the device.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely that the fluid leaked from the connection part of the plug when the clv-190 air pump is operated.Since moisture accumulates in the air system of the light source (which may have been caused by mishandling when the endoscope was cleaned manually), the system may be contaminated.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18344819
MDR Text Key330722542
Report Number3002808148-2023-14383
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
K131780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/27/2023
Initial Date FDA Received12/18/2023
Supplement Dates Manufacturer Received02/01/2024
Supplement Dates FDA Received02/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/07/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-