DEPUY ORTHOPAEDICS INC US PINN CAN BONE SCREW 6.5MMX25MM; BONE SCREWS AND PINS : SCREWS Back to Search Results
Catalog Number 121725500
Device Problem Osseointegration Problem (3003)
Patient Problems Inadequate Osseointegration (2646); Joint Laxity (4526)
Event Date 12/04/2023
Event Type  Injury  
Manufacturer Narrative
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
Patient was revised for dislocation.Component removal re-implantation.Additional info received from sales rep 12/11/2023 -spoke with sales rep via phone to clarify the event description.The patient was revised due to global instability.During the revision the acetabular cup and femoral component were noted to be loose, no interface provided.The original implant date was (b)(6) and revision date was (b)(6) 2023.All components were revised.Doi: (b)(6) 2023.Dor: (b)(6) 2023.Affected side: right hip.
 
Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.In order to determine if a lot related issue was possible, a worldwide complaint database search was performed.A worldwide complaint database search found no additional related reports against the provided product code/lot code combination.Based on the inability to find any additional related reports against the provided product code/lot code combination it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN CAN BONE SCREW 6.5MMX25MM
Type of Device
BONE SCREWS AND PINS : SCREWS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key18348685
MDR Text Key330803847
Report Number1818910-2023-25500
Device Sequence Number1
Product Code LPH
UDI-Device Identifier10603295009900
UDI-Public10603295009900
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K983014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number121725500
Device Lot NumberD22101279
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/05/2023
Initial Date FDA Received12/18/2023
Supplement Dates Manufacturer Received12/27/2023
Supplement Dates FDA Received12/29/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/19/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ALTRX +4 10D 32IDX50OD; CORAIL REVISION STEM STD 12; DELTA CER HEAD 12/14 32MM +1; PINN SECTOR W/GRIPTION 50MM
Patient Outcome(s) Required Intervention;
Patient Age79 YR
Patient SexFemale
-
-