Model Number BML-V442QR-30
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/24/2023
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned for evaluation.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
 
Event Description
The customer reported to olympus that during the therapeutic endoscopic retrograde cholangiopancreatography, the basket component of the lithotriptor was opened, the wires were found broken and separated from the rest of the basket.No adverse effects to patient reported.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the approved final investigation.A review of device history record and historical complaints analysis was conducted and no deviations that could have caused or contributed to the reported issue was identified.Based on the investigation result, a force exceeding the resisting force was applied to the guide wire tip.This likely caused the guide wire tip to tear.However, the root cause of the problem could not be conclusively identified.The suggested event is detectable and preventable by handling the scope in accordance with the following sections of the instructions for use: "when using the guide wire, insert the instrument with its distal tip in parallel with the guide wire while holding the distal tip as shown in figure 4.Be careful not to forcibly insert the instrument with a sharp angle between the distal tip and the guide wire as shown in figure 5.This may damage the distal tip." should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE MECHANICAL LITHOTRIPTOR V
Type of Device
SINGLE USE MECHANICAL LITHOTRIPTOR
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18349546
MDR Text Key331065296
Report Number9614641-2023-01937
Device Sequence Number1
Product Code LQC
UDI-Device Identifier04953170218422
UDI-Public04953170218422
Combination Product (y/n)N
Reporter Country CodeHR
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBML-V442QR-30
Device Lot Number38K
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/24/2023
Initial Date FDA Received12/18/2023
Supplement Dates Manufacturer Received03/04/2024
Supplement Dates FDA Received03/10/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-