Model Number BML-V442QR-30 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/24/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has not been returned for evaluation.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
|
|
Event Description
|
The customer reported to olympus that during the therapeutic endoscopic retrograde cholangiopancreatography, the basket component of the lithotriptor was opened, the wires were found broken and separated from the rest of the basket.No adverse effects to patient reported.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the approved final investigation.A review of device history record and historical complaints analysis was conducted and no deviations that could have caused or contributed to the reported issue was identified.Based on the investigation result, a force exceeding the resisting force was applied to the guide wire tip.This likely caused the guide wire tip to tear.However, the root cause of the problem could not be conclusively identified.The suggested event is detectable and preventable by handling the scope in accordance with the following sections of the instructions for use: "when using the guide wire, insert the instrument with its distal tip in parallel with the guide wire while holding the distal tip as shown in figure 4.Be careful not to forcibly insert the instrument with a sharp angle between the distal tip and the guide wire as shown in figure 5.This may damage the distal tip." should additional relevant information become available, a supplemental report will be submitted.
|
|
Search Alerts/Recalls
|