Catalog Number 328421
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/20/2023
Event Type  malfunction  
Event Description
It was reported that the bd insulin syringes with bd ultra-fine¿ needle's plunger was loose.The following was translated from chinese to english: the doctor found loose piston while injecting the patient.
 
Manufacturer Narrative
E.1.(b)(6).H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that the bd insulin syringes with bd ultra-fine¿ needle's plunger was loose.The following was translated from chinese to english: the doctor found loose piston while injecting the patient.
 
Manufacturer Narrative
H.6 investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed and no non-conformances were raised in association with this type of event for this lot.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Based on the above, no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSULIN SYRINGES WITH BD ULTRA-FINE¿ NEEDLE
Type of Device
PISTON SYRINGE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer (Section G)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer Contact
jennifer suh
5859 farinon drive
san antonio, TX 78249
8448235433
MDR Report Key18350215
MDR Text Key330877486
Report Number1920898-2023-00825
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number328421
Device Lot Number2143420
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/22/2023
Initial Date FDA Received12/18/2023
Supplement Dates Manufacturer Received12/22/2023
Supplement Dates FDA Received01/07/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/23/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-