The polarx catheter was returned to boston scientific for analysis.Upon receipt at the post market quality assurance laboratory the catheter underwent visual inspection, leakage testing, functional testing, and microscope inspection.No visible blood was seen inside the balloon and no external damage was noted.The device was assessed with an isaac pressure decay testing system to determine if any potential leak paths existed that may have contributed to the outer balloon high pressure in the field.The polarx device failed outer balloon vacuum leak testing with a gross leak.The polarx device was then connected to the smartfreeze console in attempt to recreate the blood detection error seen in the field.After multiple bending, steering, inflation, and slider switch manipulation cycles, the polarx did not trigger any blood detection errors.At this point, the device was fully assembled, and had not been dissected.The polarx device had a normal operational blood detect index of 21.The polarx device was then dissected, and the outer balloon line located inside of the handle was pressurized while submerged in water to locate a potential leak path.A leak path was found in the outer balloon itself.A small hole was found in the balloon which allowed fluid ingress, triggering a true blood detection error.The reported clinical observations were confirmed by the analysis.
|