• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION VERCISE CARTESIA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION VERCISE CARTESIA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS Back to Search Results
Model Number DB-2202-45
Device Problems High impedance (1291); Unexpected Therapeutic Results (1631); Migration (4003)
Patient Problem Insufficient Information (4580)
Event Date 11/01/2023
Event Type  Injury  
Manufacturer Narrative
Block b3: exact date unknown, event occurred in early (b)(6) 2023.Additional suspect medical device component involved in the event: product family: dbs-linear leads.Upn: m365db2202450.Model: db-2202-45.Serial: (b)(6).Batch: 7100229.
 
Event Description
It was reported the patient experienced a loss to therapy and a return of their parkinson's symptoms.High impedance measurements were observed on the lead, and x-ray imaging taken in the field revealed the lead was likely pulling out of the extension header.The physician assessed the set screw had not been fully tightened.The patient underwent a revision procedure to resecure the set screw on the extension connector to the lead.The patient did well postoperatively.Physical analysis could not be completed, as all devices remained implanted in the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERCISE CARTESIA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
erik sherburne
25155 rye canyon loop
valencia, CA 91355
7632920920
MDR Report Key18353866
MDR Text Key330864617
Report Number3006630150-2023-07926
Device Sequence Number1
Product Code NHL
UDI-Device Identifier08714729905288
UDI-Public08714729905288
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDB-2202-45
Device Catalogue NumberDB-2202-45
Device Lot Number7100129
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/20/2023
Initial Date FDA Received12/19/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/15/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age65 YR
Patient SexFemale
-
-