The customer reported to olympus, the, hf-cable had an electric arc between two hf-cable attachment pins from resectoscope during a prostate operation.The doctor was not affected, and the equipment was changed.They were using turis-electrode.The patient was not harmed but it took quite long time to change all equipment.The prostate procedure was finished successfully with the new set of devices.The procedure was extended for 20 minutes with the patient under anesthesia.The patient did not develop a permanent disability or permanent damage that caused substantial disruption in their ability to conduct normal life functions, and no additional medical or surgical intervention was required to preclude permanent impairment of a body function or damage to a body structure that is associated with the use of the olympus device.This report is related to the following linked patient identifiers: (b)(6) and (b)(6).
|