Catalog Number 80395 |
Device Problem
Overheating of Device (1437)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though results are not available as of this report.Evaluation results will be submitted as they become available.
|
|
Event Description
|
In this event it is reported that 30k fsi-sli-10s insert,pkd the insert get hot and has irregular water flow during use.No injury occurred from this alleged event.
|
|
Manufacturer Narrative
|
Returned qty.(b)(4), insert, 80395, 22222-00081176, jw, bul.Test method / gp005-wi02.Inductance: gauge id# 5349 due: 11/30/2024 result: 3.16.¿ meets spec ¿ does not meet spec ¿ n/a.Tip condition: - ¿ meets spec ¿ does not meet spec ¿ n/a.Stack condition ¿ meets spec ¿ does not meet spec ¿ n/a.Tested on: g136 gage id# 9626-2 due date: 03/31/24 set pressure: 35 psi.Water flow: output rate: 35 psi - ¿ meets spec ¿ does not meet spec ¿ n/a.Spray pattern: - ¿ meets spec ¿does not meet spec ¿ n/a.Water leak: ¿ hp sealing ring ¿ proximal ring ¿ distal ring.¿ seam leak ¿ n/a.Vibration: - ¿ meets spec ¿ does not meet spec ¿ n/a.Grip condition: ¿ meets spec ¿ does not meet spec ¿ n/a.Other visual observation: insert tip has something obstructing the water flow, is look little clogged.Alleged event: ¿ confirmed - ¿ not confirmed.
|
|