Model Number TABLETOP |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Loss of Vision (2139); Phototoxicity (2165); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 11/27/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A physician reported that after a vitrectomy surgery using an ophthalmic system and forceps, three patients experienced post operative problems was vision loss, inner retina hyper-reflectivity and phototoxicity.Patient symptom were continuing.Additional information has been requested.
|
|
Manufacturer Narrative
|
The company representative was unable to confirm nor replicate the reported event.The system was tested and found to meet product specifications.A non-conformance-based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this serial number was performed.Potentially relevant complaint was found (and reviewed and were later determined to not be relevant to this investigation.The root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
|
Search Alerts/Recalls
|