• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (SPARKS) BD TOTALYS SLIDEPREP; PROCESSOR, CERVICAL CYTOLOGY SLIDE, AUTOMATED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (SPARKS) BD TOTALYS SLIDEPREP; PROCESSOR, CERVICAL CYTOLOGY SLIDE, AUTOMATED Back to Search Results
Catalog Number 491346
Device Problem Contamination (1120)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/23/2023
Event Type  malfunction  
Manufacturer Narrative
H3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that during use with the bd totalys slideprep, there was suspected cross contamination of samples.No patient impact was reported.
 
Event Description
It was reported that during use with the bd totals slide-prep, there was suspected cross contamination of samples.No patient impact was reported.
 
Manufacturer Narrative
Complaint reports cross contamination on slide-prep (catalog number 491346) serial number (b)(6).Complaint alleges possible cross contamination; no erroneous results were recorded.Service checked for vacuum issue and quad bundle damage; no issues were detected.Service performed a water run, random xyz test, and a tip pickup test without any issues or errors.Instrument is working normally.Root cause not determined, and this complaint is not a confirmed failure of the instrument based on the service investigation.Review of device history record for instrument serial number, (b)(6) is not required because this complaint does not allege an early life failure or failure at installation and has changed configuration since release from manufacturing due to service repairs/pms.Device was installed on 01/24/2013.Service history review was performed for the instrument (b)(6) , and no additional work orders were observed for the complaint failure mode reported.Review of risk management files confirms there are no new or modified risks associated with this failure mode.H3 other text : see h.10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD TOTALYS SLIDEPREP
Type of Device
PROCESSOR, CERVICAL CYTOLOGY SLIDE, AUTOMATED
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
jo doyka
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key18355304
MDR Text Key330879427
Report Number1119779-2023-01381
Device Sequence Number1
Product Code MKQ
UDI-Device Identifier00382904913465
UDI-Public(01)00382904913465
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P970018
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number491346
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/23/2023
Initial Date FDA Received12/19/2023
Supplement Dates Manufacturer Received12/22/2023
Supplement Dates FDA Received01/18/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/19/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-