Product analysis: it was determined on analysis that the programmer tested out of specification as it was identified that the stylus was not working correctly, the connector on the cable was damaged.It was also determined at analysis that the touchscreen glass was broken in the lower left corner.The corner of the touchscreen had a large dark area and was not functional.There was no paper in the paper tray and the left and right display hasps were broken.The cord bay was broken, the right latch was broken off.The upper and lower bullets and keyboard cover plate were missing.The radiofrequency head connector on the link electronic module (lem) board was damaged.The software was reinstalled and updated to the newest version.All found defective parts were replaced and all other identified issues were resolved.The device then passed all final functional tests.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|