Model Number M00546650
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/14/2023
Event Type  malfunction  
Event Description
It was reported to boston scientific corporation that spy ds controller were used during an endoscopic retrograde cholangiopancreatography cholangioscopy procedure performed for the treatment of complex stone disease in the bile duct on (b)(6) 2023.During procedure, the controller turns off after three to four minutes of use.The procedure was not completed due to this event.There were no patient complications reported as a result of this event.
 
Manufacturer Narrative
Block h6 (impact codes): imdrf impact code f1001 is being used to capture the reportable event of aborted/cancelled procedure.
 
Manufacturer Narrative
Block h6 (impact codes): imdrf impact code f1001 is being used to capture the reportable event of aborted/cancelled procedure.Block h10: the returned spyglass ds digital controller was analyzed by enercon technologies, and a visual evaluation noted that the front panel has cosmetic damage, top cover and main housing have finish damage.The controller did not have any unexpected shutdown power problems; however, there was no video output when a catheter was inserted.This was caused by fluid ingress inside the unit housing and light engine.The reported event was not confirmed.A risk review confirms this is not a new or unanticipated event.As per report attached, the problems are unlikely related to manufacturing, as product analysis identified problems related to use and maintenance and did not identify any manufacturing defect.In addition, during a device history record review conducted by enercon, it was confirmed that the device met all manufacturing specifications.Although the number of procedures/recycles of the unit are unknown, improper handling of the device or wear/tear on internal components over time likely contributed to the event.Based on all gathered information, the conclusion code selected for this event is cause traced to maintenance, which indicates that problems are traced to improper routine or preventative maintenance.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
 
Event Description
It was reported to boston scientific corporation that spy ds controller were used during an endoscopic retrograde cholangiopancreatography cholangioscopy procedure performed for the treatment of complex stone disease in the bile duct on (b)(6) 2023.During procedure, the controller turns off after three to four minutes of use.The procedure was not completed due to this event.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPYGLASS DS
Type of Device
LED LIGHT SOURCE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
ENERCON TECHNOLOGIES
25 northbrook dr
665
gray ME 04039
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key18356344
MDR Text Key331210059
Report Number3005099803-2023-06646
Device Sequence Number1
Product Code NTN
UDI-Device Identifier08714729874348
UDI-Public08714729874348
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K183636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00546650
Device Catalogue Number4665
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/24/2023
Initial Date FDA Received12/19/2023
Supplement Dates Manufacturer Received02/06/2024
Supplement Dates FDA Received03/05/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/19/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-