• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NXSTAGE MEDICAL, INC. NXSTAGE SYSTEM ONE; HIGH PERMEABILITY HEMODIALYSIS SYSTEM Back to Search Results
Model Number NX1000-16
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Cardiac Arrest (1762); Loss of consciousness (2418)
Event Date 11/19/2023
Event Type  Death  
Event Description
A report was received on 21 nov 2023 from the home therapy nurse (htn) of a 47 year old male patient with an extensive medical history including end stage renal disease, who stated the patient went into cardiac arrest during a home hemodialysis treatment on (b)(6) 2023.Additional information was received on 27 nov 2023 from the htn, who stated the patient was about an hour into treatment when he lost consciousness and the caregiver called emergency medical services.Per the htn the patient subsequently expired at an unspecified time on (b)(6) 2023.
 
Manufacturer Narrative
The device was received for evaluation and successfully passed testing.A review of the device history record (dhr) was conducted which confirmed that the device met all quality criteria and manufacturing specifications prior to release.Available log files were retrieved and analyzed which showed device performance was without deficiency and was unremarkable.There was no indication of a device malfunction from the available information.The nxstage system one user guide outlines risks associated with performing hemodialysis therapy and warns all treatments must be administered under a physician''s prescription and must be performed by a trained and qualified person, considered to be competent in the use of this device by the prescribing physician.Udi: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NXSTAGE SYSTEM ONE
Type of Device
HIGH PERMEABILITY HEMODIALYSIS SYSTEM
Manufacturer (Section D)
NXSTAGE MEDICAL, INC.
350 merrimack street
lawrence MA 01843
Manufacturer (Section G)
MEDIMEXICO S. DE R. L. DE C. V
av. valle imperial no. 10523
parque industrial valle sur
tijuana 22180
MX   22180
Manufacturer Contact
paula rogalski
nxstage medical, inc.
350 merrimack street
lawrence, MA 01843
9784505276
MDR Report Key18357379
MDR Text Key330896350
Report Number3003464075-2023-00112
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K170469
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 12/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberNX1000-16
Device Catalogue NumberCYCLER VERSIHD 1.0, NO NIBP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received12/19/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/02/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age47 YR
Patient SexMale
Patient Weight115 KG
-
-