H.6.Investigation summary: material 245122 is manufactured by rehydrating the media components with usp purified water, and thoroughly mixing until a homogeneous solution is obtained.The tubes are filled, capped, torqued, and then labeled by machine per standard operating procedure (sop).The tubes are terminally autoclaved in an air over pressure (aop) autoclave, per manufacturing instructions, using a validated cycle.Post autoclaving, tubes are packaged into final shipping configurations.The batch history record review for batch 1276310 was satisfactory per internal procedures.Formulation, filling, torquing, packaging processes were within specifications.In process checks were performed during manufacturing at designated intervals per procedures.Checks for fill volume were complete and within specifications per procedures and checks for torque confirmed that the caps were tightened to the validated specifications per internal procedure.Qc inspection and testing were satisfactory at time of release.The complaint history was reviewed, and no other complaints have been taken on this batch.Retention samples from batch 1273610 were not available for inspection.Three (3) photos were received for review for this complaint.The first photo shows a single tube held above a tray of tubes with batch 1273610 and expiration 2023-03-26 in view.There does not appear to be any defect with this tube.The second photo shows a single tube with a low fill held above a tray of tubes.The third photo shows a single tube held above a tray of tubes; the tube has no label.No returns were received for investigation.The complaint can be confirmed.No complaint trends for this defect have been identified for this product; no actions are identified at this time.Bd will continue to trend complaints for defects.
|