BECTON, DICKINSON & CO. (SPARKS) BD BBL¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, 7ML; SYSTEM, BLOOD CULTURING Back to Search Results
Catalog Number 245122
Device Problem Device Markings/Labelling Problem (2911)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/23/2023
Event Type  malfunction  
Manufacturer Narrative
H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported when using the bd bbl¿ mgit¿ mycobacteria growth indicator tubes, 7ml had missing labels.There was no report of impact to patient or user.
 
Manufacturer Narrative
H.6.Investigation summary: material 245122 is manufactured by rehydrating the media components with usp purified water, and thoroughly mixing until a homogeneous solution is obtained.The tubes are filled, capped, torqued, and then labeled by machine per standard operating procedure (sop).The tubes are terminally autoclaved in an air over pressure (aop) autoclave, per manufacturing instructions, using a validated cycle.Post autoclaving, tubes are packaged into final shipping configurations.The batch history record review for batch 1276310 was satisfactory per internal procedures.Formulation, filling, torquing, packaging processes were within specifications.In process checks were performed during manufacturing at designated intervals per procedures.Checks for fill volume were complete and within specifications per procedures and checks for torque confirmed that the caps were tightened to the validated specifications per internal procedure.Qc inspection and testing were satisfactory at time of release.The complaint history was reviewed, and no other complaints have been taken on this batch.Retention samples from batch 1273610 were not available for inspection.Three (3) photos were received for review for this complaint.The first photo shows a single tube held above a tray of tubes with batch 1273610 and expiration 2023-03-26 in view.There does not appear to be any defect with this tube.The second photo shows a single tube with a low fill held above a tray of tubes.The third photo shows a single tube held above a tray of tubes; the tube has no label.No returns were received for investigation.The complaint can be confirmed.No complaint trends for this defect have been identified for this product; no actions are identified at this time.Bd will continue to trend complaints for defects.
 
Event Description
It was reported when using the bd bbl¿ mgit¿ mycobacteria growth indicator tubes, 7ml had missing labels.There was no report of impact to patient or user.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BBL¿ MGIT¿ MYCOBACTERIA GROWTH INDICATOR TUBES, 7ML
Type of Device
SYSTEM, BLOOD CULTURING
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
jo doyka
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key18357406
MDR Text Key331241412
Report Number1119779-2023-01383
Device Sequence Number1
Product Code MDB
UDI-Device Identifier30382902451229
UDI-Public(01)30382902451229
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K974883
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date03/26/2023
Device Catalogue Number245122
Device Lot Number1273610
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/23/2023
Initial Date FDA Received12/19/2023
Supplement Dates Manufacturer Received01/18/2024
Supplement Dates FDA Received02/02/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/30/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-