Catalog Number UNK HIP ACETABULAR CUP DURALOC |
Device Problem
Osseointegration Problem (3003)
|
Patient Problem
Inadequate Osseointegration (2646)
|
Event Date 12/05/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Product complaint # (b)(4).D4: the device catalog number is unknown; therefore, udi is unavailable.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
Patient was revised due to loosening of cup at the unknown interface.The patient had a tha with an srom stem and duraloc cup.He returned to surgery due to poly wear, for a head and liner exchange.When trying to removing the liner it was noted that the cup was loose.It was removed.A new pinnacle cup and liner was placed.No item lot numbers for the removed implants or the retained implants are available.There were no deficiencies, delays or broken instruments.Doi: 2002.Dor: (b)(6), 2023.Affected side: left.
|
|
Event Description
|
Additional information was received and stated that there were no delays or adverse events.The loosening was in the implant bone interface.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.An evaluation of the manufacturing record could not be performed as the required product/lot number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the device lot number is unknown, therefore a device history review could not be performed. if the lot/serial number becomes available, the record will be re-assessed.
|
|