At the start of the atrial fibrillation case, impedance errors resulted in a procedural delay.When disconnecting a lead from the module, the issue resolved and the procedure was completed with no adverse consequences to the patient.A patch impedance error was noted on the system and there was no ecg (purple lines).A new system reference patch was tried, ecg gel was applied to the patch, and the patient was re-patched, but the issue persisted.When disconnecting the right leg lead from the surfacelink, the signals returned on ensite x.The procedure was delayed, and the physician felt that it was clinically significant.The procedure was completed without adverse consequences for the patient.
|
One ensitex surfacelink (surflink) module was received for evaluation.Based on the investigation and information provided to abbott, the reported event was confirmed, and the root cause was attributed to an electrical open to the active belly patch pin from an undetermined event.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined that no non-conformances associated with the reported event were identified.
|