• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WOLVERINE CORONARY CUTTING BALLOON MONORAIL; CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WOLVERINE CORONARY CUTTING BALLOON MONORAIL; CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING Back to Search Results
Model Number 3851
Device Problem Difficult to Advance (2920)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/22/2023
Event Type  malfunction  
Manufacturer Narrative
E1 - initial reporter facility name: (b)(6) hospital.Device evaluated by mfr: the device was returned for evaluation.Visual, tactile, microscopic and device-to-device interaction analysis was performed on the device.A visual examination of the balloon identified no damages.Balloon folds were tightly folded.No issues were noted along the hypotube shaft.A cut was noted on the outer polymer extrusion, 5mm cut from port exchange.A detailed microscopic examination of the balloon material identified no tears or pinholes in the balloon.All blades were fully bonded on the balloon and did not exhibit any signs of damage.Tip showed no signs of tip damage.A microscopic examination of the proximal and distal markerbands identified no damage.The device could be loaded and tracked on a 0.0140" test guidewire and a 5f guide catheter with no issues.No other issues were identified during the product analysis.
 
Event Description
Reportable based on device analysis completed on (b)(6) 2023.It was reported that there was resistance to advance through the guide catheter.The 38 mm x 2.5 mm, 75% stenosed target lesion was located in the moderately tortuous and severely calcified left anterior descending artery (lad).A 10mm x 2.25mm wolverine coronary cutting balloon was selected for use.During the procedure, it was noted that after the tip was connected to the guide wire, there was resistance in the internal delivery of the guide catheter, and it was difficult to smoothly advance through the guide catheter.There a significant bend involved in the target lesion.The procedure was completed with another of the same device.There were no patient complications reported, and patient was stable post procedure.However, device analysis revealed that there was a cut was noted on the outer polymer extrusion, 5mm cut from port exchange.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WOLVERINE CORONARY CUTTING BALLOON MONORAIL
Type of Device
CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18358596
MDR Text Key330904851
Report Number2124215-2023-68412
Device Sequence Number1
Product Code NWX
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 12/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3851
Device Catalogue Number3851
Device Lot Number0031260738
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/30/2023
Initial Date FDA Received12/19/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/20/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age49 YR
Patient SexFemale
Patient Weight50 KG
-
-