• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ELEKTA SOLUTIONS AB MOSAIQ; ACCELERATOR, LINEAR, MEDICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ELEKTA SOLUTIONS AB MOSAIQ; ACCELERATOR, LINEAR, MEDICAL Back to Search Results
Device Problem Computer Software Problem (1112)
Patient Problem Insufficient Information (4580)
Event Date 12/01/2023
Event Type  malfunction  
Manufacturer Narrative
The manufacturer's investigation is on-going and further information will be provided once the investigation has completed.
 
Event Description
The customer reported that a portal image was acquired that indicted that some multi leaf collimators (mlcs) were not in the correct position.
 
Manufacturer Narrative
Section h6 updated section h11 updated the investigation was completed by conducting a thorough evaluation of the product and the reported information.The shape that was seen for the field in the screenshot is the completely irradiated area outline (ciao) of the field.The field is a stepnshoot type field where collimation changes during beam-on.The mlc correctly moved to the ciao position for the planned port exposure.Based upon the available information there was no patient mistreatment.Mosaiq did not have any malfunction and was working as designed and intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOSAIQ
Type of Device
ACCELERATOR, LINEAR, MEDICAL
Manufacturer (Section D)
ELEKTA SOLUTIONS AB
kungstensgatan 18
box 7593
stockholm, SE-10 3 93
SW  SE-103 93
Manufacturer (Section G)
ELEKTA SOLUTIONS AB
400 perimeter center terrace
suite 50
atlanta GA 30346
Manufacturer Contact
cornerstone
london road
crawely, west sussex RH10 -9BL
MDR Report Key18362311
MDR Text Key331338916
Report Number3015232217-2023-00072
Device Sequence Number1
Product Code IYE
UDI-Device Identifier00858164002367
UDI-Public(01)00858164002367(10)2.83.053
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K183034
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/04/2023
Initial Date FDA Received12/20/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/31/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-