Catalog Number 04625412003 |
Device Problems
Contamination (1120); Display Difficult to Read (1181)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/08/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Section e3: occupation is patient/consumer.The meter was requested for investigation.
|
|
Event Description
|
We received an allegation of a display issue with a coaguchek xs meter.The customer stated the display was faded and "some" of the segments from the results field were missing.
|
|
Manufacturer Narrative
|
The meter was received for investigation.The circuit board showed contamination caused by penetrated liquid.Some of the segments of the device displayed with a weak contrast.The contamination observed affects the conductive rubber contacts and is due to improper handling or maintenance.
|
|