Model Number URF-P7 |
Device Problem
Peeled/Delaminated (1454)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
The customer reported to olympus, the uretero-reno fiberscope had broken fiber.The issue was found during an unknown event.The device was returned for evaluation.During the device evaluation, peeled coating on the connecting tube was found.There were no reports of patient harm.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
|
|
Manufacturer Narrative
|
The device was returned to olympus for evaluation and the evaluation found significant breakages on the image guide bundle, water tightness lost due to a dented scope cover, a deformed forceps lever, and the bending angle in the up direction did not meet the standard value due to the wear of angle wire.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the peeled connecting tube was stress of repeated use, external factors, or handling.Olympus will continue to monitor field performance for this device.
|
|