• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. VISERA 4K UHD XENON LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. VISERA 4K UHD XENON LIGHT SOURCE Back to Search Results
Model Number CLV-S400
Device Problems No Device Output (1435); Connection Problem (2900); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned as part of the asset return process, in addition to the reportable findings in b5, the following findings were discovered; the front panel is crushed and the top cover was deformed.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
 
Event Description
A user facility returned the olympus asset, xenon light source, to the olympus service center.Upon inspection and testing of the returned device, it was observed that due to deformation of the scope socket, the endoscope could not be identified, the base did not hold the light guide cable, and the high function was not available.This report is being submitted for the event found during evaluation.There was no patient involvement.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over eight (8) years since the subject device was manufactured.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely that the light cable is not recognized due to the deformed scope socket.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA 4K UHD XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18367195
MDR Text Key331305491
Report Number3002808148-2023-14558
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170374579
UDI-Public04953170374579
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K151011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-S400
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/12/2023
Initial Date FDA Received12/20/2023
Supplement Dates Manufacturer Received01/29/2024
Supplement Dates FDA Received02/12/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/05/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-