Manufacturer's investigation conclusion: the reported event of the centrimag console battery not passing the preventative maintenance test after two attempts was not confirmed.The centrimag console (serial number (b)(6)) was not returned for analysis.Additionally, there were no log files, photos, or any other supporting documents submitted that would confirm the reported event.It was further reported that the battery was installed in (b)(6) 2022, and the expiration date was in 2025; therefore, the battery was not expired at the time of the event.Provided information stated that the serial number of the console is (b)(6), the serial number of the battery is (b)(6), and the battery is not returning.The root cause of the reported event could not be conclusively determined through this analysis.The device history records were reviewed for the centrimag 2nd generation primary console (serial number: (b)(6)) and the console was found to pass all manufacturing and quality assurance specifications.The 2nd generation centrimag system operating manual section 4 entitled "warnings & precautions" warns "one additional 2nd generation centrimag primary console, motor and flow probe are required as backup system in the immediate vicinity of each patient whenever the centrimag or pedivas blood pump is used.The backup console must be connected to the backup motor and to the backup flow probe, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the main console, motor or flow probe experience a malfunction." the 2nd generation centrimag system operating manual section 10 entitled "emergency and troubleshooting" states that "the recommended practice whenever there is a 2nd generation centrimag primary console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the backup motor and console to continue patient support.Do not exchange individual motors or individual consoles during patient support." the 2nd generation centrimag system operating manual table 15 entitled ¿console maintenance schedule¿ addresses how often to perform maintenance including when to perform battery maintenance and when to replace the battery.No further information was provided.The manufacturer is closing the file on this event.
|